4000 X-Series Oscilloscopes Advanced Training Guide
4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
13
Set the scope timebase to
2.000 ms/div
.
14
Press the
[Search]
feature in the Horizontal area of the front panel.
15
You will notice a display pop up telling you to enable the
Lister
feature in order
to search. Do this by tapping the double down arrows in the top right corner of
the waveform display, which brings up a half screen
Lister
window. This is a list
of all decoded words and errors.
16
Ensure that
Search for
is set to
Label
, and then tap the
Label
softkey. Use the
Entry knob to set a value of
076
.
White arrows should appear on the top of the waveform, below the Lister window.
These represent the locations of all hits from the search parameters, namely all
words with a label of 076. There should be 10-11 on the display. They are also
marked in the Lister window with orange tabs on the left column.
17
Press
[Run/Stop]
to halt all acquisitions.
18
Use the
and
front panel navigation keys to automatically scroll through
all search results. The lower screen will automatically zoom into the word in the
search list, and point to it in the lister window. See
.
Figure 96
Triggering on a frame with label 076.