72
4000 X-Series Oscilloscopes Advanced Training Guide
3
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
Your scope’s display should now look similar to
while using the scope’s
default trigger condition (rising edge of channel-1). Let’s now set up the scope to
trigger on one of the wider digital pulses on channel-2. To do this we will qualify
triggering on the channel-1 pulse, then delay arming of the trigger past the 5 MHz
analog burst, and then finally trigger on the Nth occurrence of one of the wider
digital pulses.
12
Press the
[Trigger]
front panel key; then select
Edge then Edge
.
13
Double-tap the
Delay
softkey; then enter
2.0 µs
.
14
Double-tap the
Nth Edge B
softkey and enter
3
.
Figure 44
Capturing a complex burst signal while triggering on a synchronization signal.