58
4000 X-Series Oscilloscopes Advanced Training Guide
3
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
15
Tap the “
< >
” softkey to select “
>
”.
16
Tap the
Time
softkey twice, which opens the keypad. Set the violation time to
>
100 ns
.
Your scope should now be triggering on just rising edges of the input signal that
have rise times in excess of 100 ns as shown in
. In addition to triggering
on Rise/Fall Time violation edges, the Keysight 4000 X-Series oscilloscopes can
also perform
Search & Navigation
to find multiple edge speed violation conditions,
regardless of the specific trigger setup condition.
17
Set the scope’s timebase to
200.0 µs/div
.
18
Press
[Run/Stop]
to stop repetitive acquisitions (the
[Run/Stop]
key should turn
red).
19
Press the
[Search]
front panel key.
20
Tap the
Search Edge
softkey and select the
Rise/Fall Time
search mode.
21
Tap the
Settings
softkey.
22
Tap the “
< >
” softkey; the select “
>
”.
23
Press the
Time
softkey; enter in
> 100 ns
.
Figure 33
Using Rise/Fall Time triggering to trigger on violation edge speeds.