Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
4
4000 X-Series Oscilloscopes Advanced Training Guide
89
17
Tap the
MOSI
softkey; then select
D9
.
18
Press the (Back) front panel button to return to the previous menu.
19
Tap the
MISO
softkey; then select
D8
.
20
Press the (Back) front panel button to return to the previous menu.
21
Tap the
CS
softkey; then tap the
~CS
softkey; then select
D6
using the Entry
knob.
22
Press the
[Label]
front panel key to turn on the scope’s default labels.
Your scope should now be decoding the MOSI and MISO serial data lines similar to
. However, the scope is not yet triggering on these signals. The scope
should be randomly auto triggering because the default trigger condition is to
trigger on a rising edge of channel-1. But there are no signals present on the
channel-1 input. Let’s now set up the scope to trigger on a MOSI data byte equal
to 0000 0011 (03
HEX
).
23
Press the
[Trigger]
front panel key; then select
Serial 1 (SPI)
.
24
Tap the
Trigger Setup
softkey.
25
Tap the
Bit
softkey and enter
0
in the keypad.
26
Tap the
0 1 X
softkey to select
0
as the trigger logic condition for Bit 0 (MSB).
Figure 56
Decoding a 4-wire SPI serial bus.
Back
Back