62
Keysight U3045AM04 User’s and Service Guide
U3045AM04
Service Information
1-
5.
Control Lines are not working.
a.
Verify that the control voltage pin connections to the DUT control
lines are connected properly. Refer to
.
b.
Verify that the rear panel DC voltage control adjustment can be set
to 5 Vdc. Refer to
.