Keysight U3045AM04 User’s and Service Guide
45
U3045AM04
Control Lines
After a power reset all DUT control lines are initially configured to a logic high
state or connected to Pin 13, refer to
. To reset all control
lines to logic high, without having to reset the power switch on the test set,
make the following analyzer entry:
Front panel analyzer Interface Control Mode line entry = 112.0 > OK.
Always determine which control lines you want set to a logic zero, the other
remaining lines will automatically be set to a logic high. Note the <data> value
of these lines and calculate the SUM of their <data> values.
Listed are two examples to illustrate this concept. Refer to
, shown
with all lines = logic high.
Example 1: To change lines 1 & 8 to equal logic Low, all others logic high.
1.
Line 1 (<address> = 112 and <data> = 1)
2.
Line 8 (<address> = 112 and <data> = 128)
3.
The SUM of the <data> values = 129
4.
Front panel analyzer Interface Control Mode line entry = 112.129 > OK.
Example 2: From Example 1 to only change Lines 2 & 3 to equal logic low, all
others logic high.
1.
Line 2 (<address> = 112 and <data> = 2)
2.
Line 3 (<address> = 112 and <data> = 4)
3.
The SUM of the <data> values = 6
4.
Front panel analyzer Interface Control Mode line entry = 112.6 > OK.
Since all control lines have the same <address>, only one "<address>.<data>"
command line is needed to control all 8 lines.