16
Keysight U3045AM04 User’s and Service Guide
U3045AM04
N5244A or N5245A RF Interface Cable Connections (U3021-60089)
1-
N5244A or N5245A RF Interface Cable Connections (U3021-60089)
illustrates the cable configuration of the test set to the network
analyzer. The cables have been supplied with Cable Kit (U3021-60089).
1.
Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM
jumpers (x8) on the network analyzer. The RCVR R1 to SOURCE OUT
reference loop jumpers (x4) remain on the front panel. Keep the jumpers
for future use.
2.
Connect the RF interconnect cables from the network analyzer to the test
set in the order listed. As you are connecting each cable, torque to 8 in-lb.
.
Over-torque will cause damage to the test set and may cause connectors to
spin or become loose.
illustrates the final two digits of the part number for each cable. The
cables must be connected in the numeric order listed in
.
Figure 10
RF Interface Connections
Table 6
RF Interface Cable Connection (U3021-60089)
Numeric
Order
RF Cables
From:
Network Analyzer
To:
Test Set
1
U3047-20016
CPLR ARM
CPLR ARM
2
U3047-20016
RCVR IN
RCVR OUT
3
U3047-20017
CPLR THRU
CPLR THRU
4
U3047-20017
SOURCE OUT
SOURCE IN
17 (x8)
16 (x8)