44
Keysight U3045AM04 User’s and Service Guide
U3045AM04
Control Lines
1-
Setting the Control Lines with Address and Data Values
This section describes how to control the rear panel DUT control lines.
The following control feature will function only while the analyzer is in
Standalone mode.
Setting the Analyzer to Standalone Mode
1.
Select Utility > System > Configure > Multiport Capability. In the dialog
box select Restart as a standalone PNA > OK.
Table 11
As described in
“Setting the Test Port Paths with Address and Data” on page
, the <address>.<data> values are determined in a similar man
ner
, with the
following exceptions:
—
will be used.
—
The <address> value is always = 112
Test Set DUT Control Address and Data Logic Table
<Address>
<Data>
Description
Line/Pin
112
0
ALL DUT Control Lines set to logic high or connected to Pin 13
1-8
112
255
ALL DUT Control Lines set to logic low or connected to Pin 14
1-8
112
1
DUT Control Line 1 set to logic low or connected to Pin 14
1
112
2
DUT Control Line 2 set to logic low or connected to Pin 14
2
112
4
DUT Control Line 3 set to logic low or connected to Pin 14
3
112
8
DUT Control Line 4 set to logic low or connected to Pin 14
4
112
16
DUT Control Line 5 set to logic low or connected to Pin 14
5
112
32
DUT Control Line 6 set to logic low or connected to Pin 14
6
112
64
DUT Control Line 7 set to logic low or connected to Pin 14
7
112
128
DUT Control Line 8 set to logic low or connected to Pin 14
8