Keysight U3045AM04 User’s and Service Guide
U3045AM04
Operational Check with Cal Kit
1-
Figure 37
Calibration Complete
7.
Repeat
Connect the ECal or the mechanical cal kit to Port 1 or
port you are testing. Torque to 8 in-lb. For further information
49 thru
Completed prompt, select Save As User Calset, type the name
Overwrite the Calset if it already exists and Save.
for
Ports
2
thru 8. When finished, there should be 8 Cal Sets saved with the
titles
“999.1” thru “999.8” (8-Port).
8.
Select Trace/Chan > Trace > Delete Trace. There should be no traces on
the analyzer’s screen.
9.
To launch the Cal Set Viewer toolbar. Select Response > Cal > Manage
Cals > Cal Set Viewer.
Figure 38
Calibration, Cal Set Viewer
10.
From the Cal Sets drop-down menu, select 999.1 and select Enable.
Select the Reflection Tracking(x,x), where x,x is the port being tested.
Ensure that the Enable and Error Terms check boxes are selected.