U3045AM04
Operational Check with Cal Kit
1-Port Calibration and Verification Procedure
1.
Connect the ECal or the mechanical cal kit to Port 1 or the port you are
testing. Torque to 8 in-lb. For further information refer to the Help menu,
see
2.
Perform a 1-Port Calibration on Port 1. On the analyzer, select Response >
Cal > Start Cal > Calibration Wizard.
a.
If using a mechanical cal kit, select SmartCal (Guided Calibration) >
Next.
b.
If using an ECal module, connect the ECal to a analyzer USB port.
Select Use Electronic Calibration (ECal) > Next.
3.
Continue following the Cal Wizard prompts. In the Select Calibration Ports
and ECal Module dialog box click Clear All, then select Port 1 > Next >
Measure.
4.
Ensure the Cal Kit you are using is indicated on the right side of the
window.
Figure 36
1-Port Calibration
5.
Continue to follow the prompts until the Calibration Completed dialog box
appears.
6.
At the Calibration Completed prompt, select Save As User Calset, type the
name 999.1. Overwrite the Calset if it already exists and Save.
If you do not have a key board, select Save As User Calset > Edit Name and
save as 999.x. X is the port number you are calibrating. Use the numeric
keypad on the analyzer's front panel to enter “999.1.”