40
Keysight U3045AM04 User’s and Service Guide
1-
Control Lines
U3045AM04
Control Lines
The 15 pin female D-Sub connector on the rear panel provides 8 latched data
lines that can be used to control your device under test (DUT). The lines can be
controlled with the multiport external test set control, or Test Set I/O
commands. Refer to
“External Test Set Control Operation” on page
and
“Setting the Control Lines with Address and Data Values” on page
. See
for DUT control line specification.
The output voltage of the lines can be from the internal adjustable voltage
source (+2 to +5 Vdc), or an external DC power supply depending on how the
connection to the control line is configured. When using an external power
supply a positive or negative voltage can be used. Refer to
and
for control line pin location and description. Refer to
Voltage Supply Configuration” on page
for configurations.
Table 9
DUT Control Specifications
Item
Specifications
Connector Type
D–Sub, 15–pin female
Max Output Line Current
100 mA (each line)
Control Line DC resistance
< 10
Ω
(each line)
Voltage Range:
Positive Input
0 to +5 V
Negative Input
–5 to 0 V
Internal Variable Voltage
+2 to +5 V