Figure 4
7
Port 3 (20 Watt)
High power isolators should be inserted between the SRC OUT and CPLR IN
front panel ports on the Test Set if you are measuring a highly reflective
device.
The isolators placement is important to ensure that the signal is attenuated
into the U3024AH10 SRC OUT isolator connections for Port 1 and Port 2. The
reverse attenuation or isolation factor of the isolator should be 20 dB or
greater.
Booster Amplifier
RF
Input
RF
Output
Coupler
Arm
20 dB
Coupler
Main
Coupler
Output
SRC OUT CPRL THRU
Port 3
Odd Ports
RCVR IN RCVR OUT
SRC IN SRC OUT
Port 1
CPLR THRU SRC OUT
CPLR ARM A IN
SOURCE OUT CPLR THRU