The Keysight U3024AH10 is a Multiport Test Set Extension designed to be configured for high
power measurements in either a 14 or 24-Port network analyzer measurement system. Two
U3024AH10 Test Sets are required for the 24-Port network analyzer system, as shown in
. The Keysight U3024AH10 has the following key features:
•
10 Test Ports (2.4 mm male connectors)
•
Low-loss mechanical RF switching
•
High-powered device measurement capability
•
N-Port calibration and full cross bar measurement capability
•
Frequency range of operation: 10 MHz - 40 GHz (Standard Opt 700)
•
Keysight PNA-X compatibility with rear panel Test Set I/O interface for operational control.
The PNA Option 551 is required. An external personal computer is not required.
The Keysight
N5224A/B, N5225A/B,
N5244
A/B, and N524
5A
/B
PNA/
PNA-X
Network Analyzers will be referred to throughout this document as the PNA-X.
The U3024AH10 will be referred to as the Test Set.
Measurement Application Notes
Low Loss DUT Measurements
Low loss trace ripple is a product of switch repeatability and calibration technique. Low
insertion loss measurements exhibiting a 0.1 dB uncertainty ripple residing on the
measurement response is not uncommon.
Switch Configuration Command Speed
When writing address and data values to the test set directly, include a 15ms wait before
issuing another address and data set. The delay of 15ms is required for the test set internal
switches to settle. Failure to add the delay could cause the test set internal switches to stick
or cause intermittent failures.
Table 1
U3024AH10 Accessories Supplied
Description
Short, Coax 2.4 mm female
85056-60021
Part Number
RF Cable, Semi-rigid, Front Panel Jumper
N5245-20155
Guard, Jumper Cable - Side
N5242-00029
Guard, Jumper Cable - Center
N5247-00019
User's and Service Guide, U3024AH10
U3024-90001
Quantity
1
16
2
2
1
Lock-Links
1600-1423
4
1
1
1
1
1. These items are attached to the front of the Test Set at the factory.
3
Keysight U3024AH10 User's and Service Guide
Summary of Contents for U3024AH10