Measuring Booster Amplifier Gain
Table 1
7
Amplifier Inputs vs. Outputs
Input From
1
Maximum
Wattage
Output Power
Port
Reference Figure
Odd Port SRC OUT
1 Watt
All Odd Ports
See
Even Port SRC OUT
2
1 Watt
All Even Ports
See
Port 1 SRC OUT
20 Watts
Port 1
See
Port 2 SRC OUT
3
1. A 2-way high power source configuration can be configured for transmission and reverse
transmission measurements. Refer to
2. The even ports SRC OUT connection is similar to the odd ports, although located on the right side of
3. The Port 2 SRC OUT connection is similar to the odd ports, although located on the right side of the
20 Watts
Port 2
Port 3 to 12 SRC OUT
20 Watts
Port 3 to 12
Additional attenuation may be required between the coupler and the power
meter to protect the power meter from damage.
1.
Connect the Test Set to the PNA-X. See
"14-Port System Setup" on page 15
2.
Determine the Test Set and PNA-X ports that will be configured for high power measurements.
Each port can be configured separately. Refer to
the block diagrams beginning on
3.
Remove the Test Set front panel jumpers for the selected high power configuration.
4.
Connect the Coupler Main to the Amplifier Output.
5.
Connect the Amplifier Input to the Odd Port SRC OUT or the port desired to be the source. The
RF source power comes from the PNA-X SRC OUT connector. Do not connect the external
Coupler Output or Coupler ARM at this time. Refer to Table 17
6.
Turn on the PNA-X and the U3024AH10 Test Set.
7.
Select Multiport mode. Refer to
8.
Select trace S11 if you are using Port 1 or Odd Ports SRC OUT. If you are using another port as
the source port, set Sxx to the port desired. For example, S33 would be Port 3.
9.
Set the Power Level to -20 [dBm]. Save this state as User Preset to avoid an over-power
condition by the factory preset power level (-5 [dBm].
10.
Turn on the booster Amplifier.
11.
Using a power meter, measure the Coupler Output power.
55
Keysight U3024AH10 User's and Service Guide
Summary of Contents for U3024AH10
Page 5: ...U3024AH10...
Page 52: ...Figure 38 Example 1 Test Set Signal Path 47 Keysight U3024AH10 User s and Service Guide...
Page 82: ...Figure 57 24 Port System Configuration 77 Keysight U3024AH10 User s and Service Guide...
Page 83: ...Figure 58 2 Way High Power Configuration 1 Watt Keysight U3024AH10 User s and Service Guide 78...
Page 105: ......