Before Applying Power
Verify that the premises electrical supply is within the range of the instrument. The instrument
has an autoranging power supply.
If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal
condition (in which all means for protection are intact) only.
The mains wiring and connectors shall be compatible with the connector used in
the premise electrical system. Failure to ensure adequate earth grounding by not
using the correct components may cause product damage and serious injury.
This product is designed for use in Installation Category II and Pollution
Degree 2.
Verify that the premise electrical voltage supply is within the range specified on
the instrument.
Ventilation Requirements
: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.
Keysight U3024AH10 User's and Service Guide
94