Environmental Requirements
Refer to the PNA-X series standard documentation for environmental requirements.
Environmental Tests
The Test Set complies with all applicable safety and regulatory requirements for the
intended location of use.
•
Pressure Altitude (Operation)
3,000 meters (~10,000 feet)
•
The instrument can safely operate in a relative humidity of 80% for temperatures to 31
degrees C, decreasing linearly to 50% relative humidity at 40 degrees C.
Equipment Heating and Cooling
If necessary, install air conditioning and heating to maintain the ambient temperature within the
appropriate range. Air conditioning capacity must be consistent with the BTU ratings given in
Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.
Required Conditions for Accuracy Enhanced Measurement
Accuracy–enhanced (error–corrected) measurements require the ambient temperature of the
PNA-X and Test Set to be maintained within ± 1 °C of the ambient temperature at calibration.
Dimensions and Space Requirements
Standard installation of the Test Set and PNA-X includes configuration and installation on a
customer provided lab bench or table top of adequate size and strength. For weight, dimensions
and space requirements, refer to the network analyzer documentation that is used to configure
the Test Set.
Table
5
Instrument Dimensions
Model
Weight
Height
Width
Depth
N52
2
4/
2
5A
/B
49.0 kg (108 lb.)
26.
61
cm
(10.5 in)
42.5
6
cm
(16.
8
in)
5
8.23
cm
(
22.9
in)
N5244/45A/B
7
Keysight U3024AH10 User's and Service Guide
U3024AH10
11.4 kg (25 lb)
22.15 cm
(8.7 in)
42.5 cm
(16.7 in)
50 cm
(20 in)
41.8 kg (92 lb.)
Summary of Contents for U3024AH10
Page 5: ...U3024AH10...
Page 52: ...Figure 38 Example 1 Test Set Signal Path 47 Keysight U3024AH10 User s and Service Guide...
Page 82: ...Figure 57 24 Port System Configuration 77 Keysight U3024AH10 User s and Service Guide...
Page 83: ...Figure 58 2 Way High Power Configuration 1 Watt Keysight U3024AH10 User s and Service Guide 78...
Page 105: ......