background image

3 Review Hardware

S9106AC-2IF, -4UP, and -42D 5G FR1 Multi-port Transceiver Solutions, Startup Guide

23

Keysight S9106AC-42D 5G FR1 Multi-port Transceiver

 on page 

47

includes:

one S9106AC Base System (PXIe chassis and
    four M9410A PXIe VXT Vector Transceivers)

one E7770A Common Interface Unit (CIU) with
    two E7770A-HB2 CIU Channel Cards, 7 to 15 GHz High IF
    one E7770A-LA2 CIU LO Card
    one E7770A-LD2 CIU LO Distribution Card

The S9106AC-42D, four-channel configuration, consists of
four transceivers providing four RF Tx channels or four RF Rx
channels, or two High IF Tx channels or two High IF Rx channels
operating in one frequency range at a time:

FR1 RF frequencies ranging from 380 MHz to 6 GHz

or High IF frequencies from 7 to 15 GHz

This system can be configured for 2x2 or 4x4 MIMO measurements
in the FR1 RF frequency range, or 2x2 MIMO measurements
in the High IF frequency range.

Summary of Contents for S9106AC-2IF

Page 1: ... Radio 5G NR infrastructure equipment S9106AC 2IF 5G FR1 Multi port Transceiver contains two Transceiver Channels 380 MHz to 6 GHz with MIMO Subsystem and 7 to 15 GHz High IF on Two Channels S9106AC 4UP 5G FR1 Multi port Transceiver contains four Transceiver Channels 380 MHz to 6 GHz with MIMO Subsystem S9106AC 42D 5G FR1 Multi port Transceiver contains four Transceiver Channels 380 MHz to 6 GHz w...

Page 2: ...marily provided to the public Accordingly Keysight provides the Software to U S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula The license set forth in the EULA represents the exclusive authority by which the U S government may use modify distribute or disclose the...

Page 3: ...d Front Panel 25 3 2 3 Connectors S9106AC 4UP on the Rugged Front Panel 26 3 2 4 Connectors S9106AC 42D on the Rugged Front Panel 27 3 2 5 Connectors M9037A PXIe Embedded Controller 28 3 2 6 Connectors S9106AC 2IF 4UP and 42D on the PXIe Chassis Rear Panel 30 3 2 7 Connectors E7770A Common Interface Unit CIU for the S9106AC 2IF and S9106AC 42D 5G FR1 Multi port Transceiver 32 3 2 7 1 Connectors E7...

Page 4: ...to Verify Operation 54 6 1 1 Connect AC Power Cords Power On Equipment 54 6 1 2 Verify X Apps Controls S9106AC 2IF RF Ports at 3 GHz 55 6 1 3 Verify X Apps Controls S9106AC 2IF High IF Ports at 10 GHz 59 6 2 S9106AC 4UP Detailed Steps to Verify Operation 62 6 2 1 Connect AC Power Cords Power On Equipment 62 6 2 2 Verify X Apps Controls S9106AC 4UP RF Ports at 3 GHz 63 6 3 S9106AC 42D Detailed Step...

Page 5: ...wo Channels S9106AC 4UP 5G FR1 Multi port Transceiver contains four Transceiver Channels 380 MHz to 6 GHz with MIMO Subsystem S9106AC 42D 5G FR1 Multi port Transceiver contains four Transceiver Channels 380 MHz to 6 GHz with MIMO Subsystem and 7 to 15 GHz High IF on Two Channels In this document This document describes the installation process to start up and prepare the Keysight S9106AC 2IF 4UP a...

Page 6: ...2 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide ...

Page 7: ...SD on page 16 Front and Rear Panel Symbols on page 17 Returning for Service on page 19 2 1 Warning Statements and Symbols The following notices are used throughout this document Familiarize yourself with each of the notices and their meaning before operating these products A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly perf...

Page 8: ...d to maintain products in a safe condition 2 2 1 Safety Compliance These products comply with the essential requirements of the European LVD Low Voltage Directive as well as current editions of the following standards dates and editions are cited in the Declaration of Conformity IEC EN 61010 1 2 2 2 Acoustic Statement European Machinery Directive Acoustic noise emission LpA 70 dB Operator position...

Page 9: ...tive conductor inside or outside of the product is likely to make the product dangerous Intentional interruption is prohibited The Keysight S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions are designed for use in Installation Category II and Pollution Degree 2 per IEC 61010 Second Edition and 664 respectively The Keysight S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutio...

Page 10: ...s maximum option dependent E7770A Common Interface Unit CIU Height 145 6 mm 5 7 in with feet installed Width 449 mm 17 7 in across handles Depth 424 mm 16 7 in across front connectors and rear feet Weight 18 1 kg 40 lbs 2 4 Handling and Lifting Use both side handles when lifting the S9106AC Base System or E7770A Common Interface Unit CIU Use a rolling cart when transporting the S9106AC Base System...

Page 11: ...he fumes to dissipate prior to energizing the instrument Keep isopropyl alcohol away from heat sparks and flame Store in a tightly closed container The isopropyl alcohol shall not be stored or left open in the area of the equipment Use isopropyl alcohol with adequate ventilation to prevent the combustion of fumes or vapors Avoid contact with eyes skin and clothing as isopropyl alcohol causes skin ...

Page 12: ...tion Degree 2 per IEC 61010 Second Edition and 664 respectively The Keysight S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions are designed for use in the following conditions For indoor use only Altitude up to 6 561 68 ft 2 000 m Operating Temperature 10 to 40 C Maximum Relative Humidity non condensing 85 RH 8 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup G...

Page 13: ...hat the maximum ambient operating temperature of the instrument s is not exceeded by cabinet installation Keysight recommends forced air convection whenever instrument s are installed in a cabinet and further recommends that the maximum operating temperature of the cabinet be reduced 10 C from the lowest of the maximum operating temperature of a single instrument If there are any concerns or speci...

Page 14: ...t disconnects the Mains circuits from the Mains supply before other parts of the Keysight S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions The front panel switches are only standby switches and are not LINE switches Alternatively externally installed switches or circuit breakers which are readily identifiable and is easily reached by the operator may be used as disconnecting devices...

Page 15: ... ground the test set properly can result in personal injury Before turning on the test set you must connect its protective earth terminals to the protective conductor of the main power cable Insert the main power cable plug into a socket outlet that has a protective earth contact only DO NOT defeat the earth grounding protection by using an extension cable power cable or auto transformer without a...

Page 16: ...ety Protection Class I Product provided with a protective earthing ground incorporated in the power cord The mains plug shall only be inserted in a socket outlet provided with a protective earth contact Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous Intentional interruption is prohibited 2 9 3 Power Strip Warning DO NOT use a p...

Page 17: ...ust be in proper working order and provide a secure electrical connection DO NOT use outlet if the power cord does not match the outlet DO NOT use outlet if it is damaged or if the voltage is outside the required range 2 9 7 Avoid Loose Connection DO NOT use outlet if the power cord makes a loose connection 2 9 8 Make Proper Connection DO NOT allow plug to bend down or become loose ...

Page 18: ...ith a 3 0 V coin cell lithium battery This battery powers theclock circuitand retains configuration memory in CMOS RAM whilethesystem is turned off To replacetheCMOS Backup Battery 1 Turn off power to thePXIechassis 2 RemovetheM9037A embedded con troller from thechassis Observeall anti static precautions 3 Locateand removethebattery Press thebattery releaselatch 1 Thebat tery should pop out 2 4 Pl...

Page 19: ...s use the three prong AC power cord supplied with these products Failure to ensure adequate earth grounding by not using these cords can cause product damage If these products are not used as specified the protection provided by the equipment could be impaired These products must be used in a normal condition in which all means for protection are intact only Install the Keysight S9106AC 2IF 4UP an...

Page 20: ...t time each day momentarily short the center and outer conductors of the cable together Personnel should be grounded with a 1 MΩ resistor isolated wrist strap before touching the center pin of any connector and before removing any assembly from the test set Be sure that all instruments are properly earth grounded to prevent build up of static charge Perform work on all components or assemblies at ...

Page 21: ...government owned mark Products showing this mark comply with all applicable UK regulations The CE marking is a registered trademark of the European Community if accompanied by a year it is the year when the design was proven It indicates that the product complies with all relevant directives The RCM mark is a registered trademark of the Australian Communications and Media Authority The Keysight em...

Page 22: ...mber in the center indicates the time period during which no hazardous or toxic substances or elements are expected to leak or deteriorate during normal use and generally reflects the expected useful life of the product Universal recycling symbol This symbol indicates compliance with the China standard GB 18455 2001 as required by the China RoHS regulations for paper fiberboard packaging The CSA m...

Page 23: ... O boxes are not acceptable as return addresses System or module serial numbers The serial number label is located on the side panel of the module The serial number can also be read from the Soft Front Panel interface after the hardware is installed Description of failure or service required 4 Use original packaging or comparable Pack the system in its original ESD bag and packing carton If the or...

Page 24: ...20 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide ...

Page 25: ...80 MHz to 6 GHz with MIMO Subsystem and 7 to 15 GHz High IF on Two Channels Before installing hardware or software review the following hardware components Hardware Components on page 22 Connector Descriptions on page 24 Connectors 10 MHz Ref In Out on the Rugged Front Panel on page 24 Connectors S9106AC 2IF on the Rugged Front Panel on page 25 Connectors S9106AC 4UP on the Rugged Front Panel on p...

Page 26: ...equency range at a time FR1 RF frequencies ranging from 380 MHz to 6 GHz or High IF frequencies from 7 to 15 GHz This system can be configured for 2x2 MIMO measurements in the FR1 RF frequency range or 2x2 MIMO measurements in the High IF frequency range Keysight S9106AC 4UP 5G FR1 Multi port Transceiver on page 45 includes one S9106AC Base System PXIe chassis and four M9410A PXIe VXT Vector Trans...

Page 27: ...Hz High IF one E7770A LA2 CIU LO Card one E7770A LD2 CIU LO Distribution Card The S9106AC 42D four channel configuration consists of four transceivers providing four RF Tx channels or four RF Rx channels or two High IF Tx channels or two High IF Rx channels operating in one frequency range at a time FR1 RF frequencies ranging from 380 MHz to 6 GHz or High IF frequencies from 7 to 15 GHz This syste...

Page 28: ...A PXIe Reference Ref In and locks to another reference with a value of 10 MHz or 100 MHz Connector BNC f Frequency1 10 MHzor 100 MHz sinewave Lock range 1 ppm nominal Amplitude 0 to 10 dBm nominal 1The 10 MHz Ref In connector frequency range on S910xA systems is different from the M9300A PXIe Frequency Reference Data Sheet The S910xA systems only support 10 MHz or 100 MHz inputs 24 S9106AC 2IF 4UP...

Page 29: ...DC 27 dBm Maximum SafeInputPower HD Port Connector Type N f Frequency Range 380 MHzto 6 GHz Amplitude 30 VDC 27 dBm Maximum SafeInputPower RF Transceiver 7 15 GHz RF Out Connector Type N f 50 Ω nominal Frequency Range 7 to 15 GHz Amplitude 30 VDC 28 dBm Maximum ReverseInputPower RF In Connector Type N f 50 Ω nominal Frequency Range 7 to 15 GHz Amplitude 30 VDC 28 dBm Maximum SafeInputPower CIU RF ...

Page 30: ...l Frequency Range 380 MHzto 6 GHz Amplitude 30 VDC 27 dBm Maximum SafeInputPower HD Port Connector Type N f Frequency Range 380 MHzto 6 GHz Amplitude 30 VDC 27 dBm Maximum SafeInputPower Trig 1 and Trig 2 Input or Output Selectable Connectors BNC f InputImpedance InputLevel Range 1 kΩ or 50 Ω nominal 3 3 V to 3 3 V OutputImpedance OutputLevel Range 50 Ω nominal 3 3 V LVTTL 26 S9106AC 2IF 4UP and 4...

Page 31: ...DC 27 dBm Maximum SafeInputPower HD Port Connector Type N f Frequency Range 380 MHzto 6 GHz Amplitude 30 VDC 27 dBm Maximum SafeInputPower RF Transceiver 7 15 GHz RF Out Connector Type N f 50 Ω nominal Frequency Range 7 to 15 GHz Amplitude 30 VDC 28 dBm Maximum ReverseInputPower RF In Connector Type N f 50 Ω nominal Frequency Range 7 to 15 GHz Amplitude 30 VDC 28 dBm Maximum SafeInputPower CIU RF ...

Page 32: ...USB 3 0 LAN 1 and LAN 2 Connectors Two 10 100 1000BASE T RJ 45 GigabitEthernetports LANConnector Top LED Amber LANConnector Bottom LED Off Bottom LED Green Bottom LED Amber LANConnectors Top and Bottom LEDs Off Active blinks if accessing IO LANSpeed 10 Mbps LANSpeed 100 Mbps LANSpeed 1000 Mbps Network link notestablished or system is powered off LANRST Reset Button Reserved for Keysightuseonly Tri...

Page 33: ...to 4 years of service Keysight s M9037A is equipped with a 3 0 V coin cell lithium battery This battery powers the clock circuitand retains configuration memory in CMOS RAM whilethesystem is turned off To replacetheCMOS Backup Battery 1 Turn off power to thePXIechassis 2 RemovetheM9037A embedded con troller from thechassis Observeall anti static precautions 3 Locateand removethebattery Press theba...

Page 34: ...ote inhibit and power rail monitoring FAN Switch INHIBIT Switch The FAN switch controls the fan speed HIGH the fan voltage duty cycle is set to 100 which generates the highest fan speeds and best chassis cooling as well as the most fan noise AUTO the speed of the three fans are controlled based on the temperature of the chassis and the setting on the Soft Front Panel SFP Temperature parameter See ...

Page 35: ...ar Panel BNC f 50 Ω nominal Frequency Input InputSignal Inputsignal PXI timing slotPXI_CLK10_IN 10 MHz 100 PPM 100 mVPP to 5 VPP square waveor sine wave 5 V or 3 3 V TTLsignal These 10 MHz REF IN and OUT connectors are for supplying the PXI_CLK10_IN and OUT signals from the System Timing Module To provide a 10 MHz Clock to the S9106AC see Connectors 10 MHz Ref In Out on the Rugged Front Panel on p...

Page 36: ...onverting 7 to 15 GHz signals from the M9410A PXIe VXT Status LED Locked Green On indicates LO Card is locked to the10 MHzIn If LED is Red LO Card is notlocked to the10 MHzIn 10 MHz In Connector Frequency BNC f 50 Ω nominal 10 MHzreferencesignal input LO Aux Out Connector SMA f 50 Ω nominal LO Card LO Aux Out to LO Distribution Card LO In This connection supplies thelocal oscillator signal to Chan...

Page 37: ...rd 2 LO In LO Out 3 Connector SMA f 50 Ω nominal LO Distribution Card LO Out3 to Channel Card 3 LO In LO Out 4 Connector SMA f 50 Ω nominal LO Distribution Card LO Out4 to Channel Card 4 LO In LO Aux In Connector SMA f 50 Ω nominal Auxiliary inputsourcefrom theLO Card LO Outused for configuring two differentlocal oscillator sources LO Aux 1 Connector SMA f 50 Ω nominal Auxiliary local oscillator o...

Page 38: ...DUT IF In Out A Connector SMA f 50 Ω nominal In Outconnection for Channel A IF Out A Connector Type N f 50 Ω nominal ULIFoutputfor Channel A LO In Connector SMA f 50 Ω nominal Local oscillator inputfor theChannel Card IF In B Connector Type N f 50 Ω nominal DLIFinputfor Channel B DUT IF In Out B Connector SMA f 50 Ω nominal In Outconnection for Channel B IF Out B Connector Type N f 50 Ω nominal UL...

Page 39: ...ctions for the CONTROLLER CARD and all Channel Cards The rear panel of the E7770A CIU is also labeled with all required safety and compliance indicators and contains the Mains switch where the AC power cable is connected Since the E7770A LA2 CIU LO Card and the E7770A LD2 CIU LO Distribution Card do not extend through to the rear panel there are no rear connections for these two cards ...

Page 40: ...ear panel connectors for CHANNEL 1A and CHANNEL 1B are usable with the E7770A HB2 CIU Channel Card 7 to 15 GHz High IF LO CTRL PWR Not used with S9106AC 2IF S9106AC 4UP and S9106AC 42D Connector TNC f 50 Ω nominal Local oscillator control and power from LO CTRL PWR Frequency LO Power DC Power 7 to 15 GHz 10 dBm minimum 36 VDC 1A DUT IF IN HEAD IN Connector Type N f 50 Ω nominal Outputfor connectin...

Page 41: ...ed okay USB Connector Micro USB B Reserved for Keysightinternal useonly CONTROL LAN Connector 10 100 1000BASE T RJ 45 GigabitEthernetPort TRIG 1 Connector BNC f 50 Ω nominal InputLevel Range 5 V TTL Trigger signal below0 8 volts is a zero low and anything above 2 4 volts is a one high TRIG 2 Connector BNC f 50 Ω nominal InputLevel Range 5 V TTL Seedescription for TRIG 1 TRIG 3 Connector BNC f 50 Ω...

Page 42: ...38 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide ...

Page 43: ... providing two RF Tx channels or two RF Rx channels or two High IF Tx channels or two High IF Rx channels operating in one frequency range at a time FR1 RF frequencies ranging from 380 MHz to 6 GHz or High IF frequencies from 7 to 15 GHz This system can be configured for 2x2 MIMO measurements in the FR1 RF frequency range or 2x2 MIMO measurements in the High IF frequency range Keysight S9106AC 4UP...

Page 44: ... of four transceivers providing four Tx channels and four Rx channels All four channels can operate in the RF frequency range Two of the four channels can also be used in the High IF frequency range with each channel operating in one frequency range at a time FR1 RF frequencies ranging from 380 MHz to 6 GHz or High IF frequencies ranging from 7 to 15 GHz This system can be configured for 2x2 or 4x...

Page 45: ...4 Install Hardware S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide 41 4 2 Keysight S9106AC 2IF 5G FR1 Multi port Transceiver ...

Page 46: ...top right 12 13 14 15 4 S9100 60012 RF Cable Type N m to Type N m 457 mm 18 in Transceiver 1 CIU DUT Interconnections Connect CIU DUT Intcon IF Out to E7770A HB2 CIU Channel Card IF In A top left Connect CIU DUT Intcon IF In to E7770A HB2 CIU Channel Card IF Out B top left Transceiver 2 CIU DUT Interconnections Connect CIU DUT Intcon IF Out to E7770A HB2 CIU Channel Card IF In A top right Connect ...

Page 47: ...ed to verify the High IF ports as well as the HD ports see Verify Operation on page 53 RF loop back cable with adapters or equivalent can be used one 5063 1530 RF Cable SMA m to SMA m 152 4 mm 6 0 in two 1250 1250 Adapters SMA f to Type N m Remove this loop back cable from all RF ports High IF ports and HD ports during normal use 4 2 1 Connect Cables LAN Cables a Display a Keyboard Mouse 1 Connect...

Page 48: ...7770A CIU Rear Mains Switch located on the rear panel 2 Power on the E7770A CIU AC Power Switch located on the left front panel 3 Power on the S9106AC 2IF AC Power Switch located on the right front panel 44 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide ...

Page 49: ...106AC 4UP During verification only connect a loop back cable from RF OUT to RF IN on Transceiver 1 and then on Transceiver 2 Transceiver 3 and Transceiver 4 This same loop back cable with adapters can be moved to verify the HD ports see Verify Operation on page 53 RF loop back cable with adapters or equivalent can be used one 5063 1530 RF Cable SMA m to SMA m 152 4 mm 6 0 in two 1250 1250 Adapters...

Page 50: ...p above 2 Connect LAN 1 top to the Site LAN at your facility 3 Connect a Display Keyboard Mouse Not included with S9106AC 4UP 4 3 2 Connect AC Power Cord Power On Equipment Power on the S9106AC 4UP AC Power Switch located on the right front panel 46 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide ...

Page 51: ...4 Install Hardware S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide 47 4 4 Keysight S9106AC 42D 5G FR1 Multi port Transceiver ...

Page 52: ...3 14 15 4 S9100 60012 RF Cable Type N m to Type N m 457 mm 18 in Transceiver 1 CIU DUT Interconnections Connect CIU DUT Intcon IF Out to E7770A HB2 CIU Channel Card IF In A top left Connect CIU DUT Intcon IF In to E7770A HB2 CIU Channel Card IF Out B top left Transceiver 2 CIU DUT Interconnections Connect CIU DUT Intcon IF Out to E7770A HB2 CIU Channel Card IF In A top right Connect CIU DUT Intcon...

Page 53: ... be moved to verify the High IF ports on Transceiver 1 and 2 as well as the HD ports see Verify Operation on page 53 RF loop back cable with adapters or equivalent can be used 5063 1530 RF Cable SMA m to SMA m 152 4 mm 6 0 in with two 1250 1250 Adapters SMA f to Type N m 4 4 1 Connect Cables LAN Cables a Display a Keyboard Mouse 1 Connect cables as shown in the equipment setup above 2 Connect LAN ...

Page 54: ...7770A CIU Rear Mains Switch located on the rear panel 2 Power on the E7770A CIU AC Power Switch located on the left front panel 3 Power on the S9106AC 42D AC Power Switch located on the right front panel 50 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide ...

Page 55: ... port Transceiver Solutions Startup Guide 51 5 Install Software Each S9106AC 2IF S9106AC 4UP and S9106AC 42D standard configuration includes an M9037A PXIe Embedded Controller all required software comes pre installed Proceed to Verify Operation on page 53 ...

Page 56: ...52 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide ...

Page 57: ...Transceiver 2 is the next numbered slot On S9106AC 4UP and S9106AC 42D systems Transceiver 3 is the next numbered slot and Transceiver 4 is the final numbered slot S9106AC 2IF To verify operation of the S9106AC 2IF 5G FR1 Multi port Transceiver 1 Connect AC Power Cords Power On Equipment on page 54 2 Verify X Apps Controls S9106AC 2IF RF Ports at 3 GHz on page 55 3 Verify X Apps Controls S9106AC 2...

Page 58: ...el 2 Power on the E7770A CIU Rear Mains Switch located on the right rear panel 3 Power on the E7770A CIU AC Power Switch located on the left front panel 4 Connect an AC Power Cord to the S9106AC Base System right rear panel 5 Power on the S9106AC 2IF AC Power Switch located on the left front panel The M9037A PXIe Embedded Controller and the M9300A PXIe Reference are part of the S9106AC Base System...

Page 59: ...elled on the Rugged Front Panel as Transceiver 1 and Transceiver 2 6 1 2 Verify X Apps Controls S9106AC 2IF RF Ports at 3 GHz This process verifies operation of RF Transceiver 380 6 GHz RF Out and RF In located on the Rugged Front Panel 1 On RF Transceiver 1 connect an RF loop back cable from RF Out to RF In located on the Rugged Front Panel This loop back cable is not included with the S9106AC 2I...

Page 60: ...ent and has its own collection of measurement capabilities controls windows and SCPI commands Each Mode runs within a screen there can be multiple screens Screens are shown as tabs across the top of the interface IQ Analyzer Mode is included with the S9106AC 2IF and appears as a tab with the label IQ Analyzer 1 3 Select the IQ Analyzer 1 drop down selector 4 Set the Mode Measurement View for IQ An...

Page 61: ...and select RF Input b Select Frequency from the drop down menu panel top right corner select Center Frequency entry box and set it to 3 GHz c Leave all of the other RF Input settings in their preset state 6 Set the RF Output to 3 GHz this is RF Transceiver 380 6 GHz RF Out located on the Rugged Front Panel a Select Input Output from the drop down menu panel top right corner select RF Source tab se...

Page 62: ...there is a signal at 3 GHz of approximately 10 dBm RF Transceiver 380 6 GHz RF Out and RF In of the S9106AC 2IF are working properly The power level is only approximate because there is loss in the cable being used and this loss is not being corrected during the measurement 58 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide ...

Page 63: ...F In located on the Rugged Front Panel these connectors may be referred to as High IF or DUT IF 1 On RF Transceiver 1 connect an RF loop back cable from RF Out to RF In located on the Rugged Front Panel This loop back cable is not included with the S9106AC 2IF 5G FR1 Multi port Transceiver Use a cable with two adapters or equivalent such as one 5063 1530 RF Cable SMA m to SMA m 152 4 mm 6 0 in and...

Page 64: ...corner select Center Frequency entry box and set it to 10 GHz c Leave all of the other RF Input settings in their preset state 3 Set the RF Output to 10 GHz this is RF Transceiver 7 15 GHz RF Out located on the Rugged Front Panel a Select Input Output from the drop down menu panel top right corner select RF Source tab select RF Output Port drop down menu and select DUT IF OUT 1 b Set RF Power to 1...

Page 65: ...and 10 dBm a Select Peak Search from the drop down menu panel top right corner If there is a signal at 10 GHz of approximately 10 dBm RF Transceiver 7 15 GHz RF Out and RF In of the S9106AC 2IF are working properly The power level is only approximate because there is loss in the cable being used and this loss is not being corrected during the measurement ...

Page 66: ...ler and the M9300A PXIe Reference are part of the S9106AC Base System they both power on and initialize automatically when the S9106AC Base System is powered on After connecting AC power cords and powering on equipment the following process must be repeated for each M9410A PXIe VXT Vector Transceiver and its selected X Apps software pair in the system the S9106AC 4UP 5G FR1 Multi port Transceiver ...

Page 67: ...Transceiver 1 connect an RF loop back cable from RF Out to RF In located on the Rugged Front Panel This loop back cable is not included with the S9106AC 4UP 5G FR1 Multi port Transceiver Use a cable with two adapters or equivalent such as one 5063 1530 RF Cable SMA m to SMA m 152 4 mm 6 0 in and two 1250 1250 Adapters SMA f to Type N m 2 If it is not already running start the Modular TRX software ...

Page 68: ...rols windows and SCPI commands Each Mode runs within a screen there can be multiple screens Screens are shown as tabs across the top of the interface IQ Analyzer Mode is included with the S9106AC 4UP and appears as a tab with the label IQ Analyzer 1 3 Select the IQ Analyzer 1 drop down selector 4 Set the Mode Measurement View for IQ Analyzer 1 a Select Mode as IQ Analyzer Basic b Select Measuremen...

Page 69: ...and select RF Input b Select Frequency from the drop down menu panel top right corner select Center Frequency entry box and set it to 3 GHz c Leave all of the other RF Input settings in their preset state 6 Set the RF Output to 3 GHz this is RF Transceiver 380 6 GHz RF Out located on the Rugged Front Panel a Select Input Output from the drop down menu panel top right corner select RF Source tab se...

Page 70: ...there is a signal at 3 GHz of approximately 10 dBm RF Transceiver 380 6 GHz RF Out and RF In of the S9106AC 4UP are working properly The power level is only approximate because there is loss in the cable being used and this loss is not being corrected during the measurement 66 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide ...

Page 71: ...ds Power On Equipment 1 Connect an AC Power Cord to the E7770A CIU right rear panel 2 Power on the E7770A CIU Rear Mains Switch located on the right rear panel 3 Power on the E7770A CIU AC Power Switch located on the left front panel 4 Connect an AC Power Cord to the S9106AC Base System right rear panel 5 Power on the S9106AC 2IF AC Power Switch located on the left front panel The M9037A PXIe Embe...

Page 72: ... 6 3 2 Verify X Apps Controls S9106AC 42D RF Ports at 3 GHz This process verifies operation of RF Transceiver 380 6 GHz RF Out and RF In located on the Rugged Front Panel 1 On RF Transceiver 1 connect an RF loop back cable from RF Out to RF In located on the Rugged Front Panel This loop back cable is not included with the S9106AC 42D 5G FR1 Multi port Transceiver Use a cable with two adapters or e...

Page 73: ...own arrow Scroll down the list and select LaunchModularTRX Each Mode looks different and has its own collection of measurement capabilities controls windows and SCPI commands Each Mode runs within a screen there can be multiple screens Screens are shown as tabs across the top of the interface IQ Analyzer Mode is included with the S9106AC 42D and appears as a tab with the label IQ Analyzer 1 3 Sele...

Page 74: ...rner select Center Frequency entry box and set it to 3 GHz c Leave all of the other RF Input settings in their preset state 6 Set the RF Output to 3 GHz this is RF Transceiver 380 6 GHz RF Out located on the Rugged Front Panel a Select Input Output from the drop down menu panel top right corner select RF Source tab select RF Output Port drop down menu and select RF Output b Set RF Power to 10 dBm ...

Page 75: ...nd 10 dBm a Select Peak Search from the drop down menu panel top right corner If there is a signal at 3 GHz of approximately 10 dBm RF Transceiver 380 6 GHz RF Out and RF In of the S9106AC 42D are working properly The power level is only approximate because there is loss in the cable being used and this loss is not being corrected during the measurement ...

Page 76: ...h IF or DUT IF 1 On RF Transceiver 1 connect an RF loop back cable from RF Out to RF In located on the Rugged Front Panel This loop back cable is not included with the S9106AC 42D 5G FR1 Multi port Transceiver Use a cable with two adapters or equivalent such as one 5063 1530 RF Cable SMA m to SMA m 152 4 mm 6 0 in and two 1250 1250 Adapters SMA f to Type N m 72 S9106AC 2IF 4UP and 42D 5G FR1 Multi...

Page 77: ...nd select DUT IF IN 1 b Select Frequency from the drop down menu panel top right corner select Center Frequency entry box and set it to 10 GHz c Leave all of the other RF Input settings in their preset state 3 Set the RF Output to 10 GHz this is RF Transceiver 7 15 GHz RF Out located on the Rugged Front Panel a Select Input Output from the drop down menu panel top right corner select RF Source tab...

Page 78: ... there is a signal at 10 GHz of approximately 10 dBm RF Transceiver 7 15 GHz RF Out and RF In of the S9106AC 42D are working properly The power level is only approximate because there is loss in the cable being used and this loss is not being corrected during the measurement 74 S9106AC 2IF 4UP and 42D 5G FR1 Multi port Transceiver Solutions Startup Guide ...

Page 79: ...pleted before performing MIMO measurements with the S9106AC 2IF S9106AC 4UP or the S9106AC 42D To set up and run MIMO on the S9106AC 2IF S9106AC 4UP or the S9106AC 42D Refer to the Keysight MIMO Measurement Guide M9410 90016 for the Keysight M9410A M9411A VXT PXIe Vector Transceiver For DUT IF MIMO measurements select the DUT IF ports ...

Page 80: ...1740A mmWave Transceiver To access the Keysight licensing website http www keysight com find licensing To receive the latest updates by email subscribe to Keysight Email Updates www keysight com find MyKeysight To get information on preventing instrument damage www keysight com find PreventingInstrumentRepair Is your product software up to date Periodically Keysight releases software updates to fi...

Reviews: