Installation Note N9030-90098
3
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Introduction
This installation note explains how to install the hardware and licenses to add the MultiTouch user
interface to N9030A PXA signal analyzers.
This installation note supports two kits:
N9030AK-MTU - MultiTouch User Interface Upgrade
N9030AK-MTP - MultiTouch User Interface Upgrade with PC6 CPU
The N9030AK-MTU should only be installed on N9030As that already have the PC6 CPU. This kit
includes a new SSD, to ensure that when a disk drive recovery is performed, the software version
recovered will be new enough to recognize the MultiTouch front panel assembly.
The N9030AK-MTP should be installed on N9030As that do not have a PC6 CPU. A PC6 CPU is
included in this kit, along with a new SSD.
Part of the upgrade process is the license transformation process, which creates new licenses to
replace those of the existing measurement applications. This is because older measurement
application licenses whose model numbers end in the letters "A" or "B" (for example, N9071A or
N9082B) will not work with the new MultiTouch user interface. The license transformation process
will replace "A" and "B" licenses with their equivalent "C" license. For example, N9071A-2FP will be
replaced with N9071C-2FP and N9082B-1FP will be replaced with N9082C-1FP.
Products Affected:
N9030A PXA Signal Analyzer
Serial Numbers:
All
To Be Performed By:
(X) Keysight Service Center
(X) Personnel Qualified by Keysight
( ) Customer
Estimated Installation Time:
Estimated Adjustment Time:
Estimated Verification Time:
1.5 Hours
3.0 Hours
4.0 Hours