Installation Note N9030-90098
49
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Final Installation
1.
Refer to
Figure 13
. Carefully slide the instrument cover back onto the instrument from the rear
of the analyzer, making sure not to damage any internal cables. The seam on the cover should
be on the bottom of the instrument. Be sure the cover seats into the gasket groove in the Front
Frame Assembly.
2.
Replace the four rear feet
(4)
to the rear of the instrument using the four screws
(3)
. Torque
to 21 inch pounds.
3.
Replace the strap handles
(2)
on both sides of the instrument using the four screws
(1)
.
Torque to 21 inch pounds.
4.
Locate the Top Trim Strip, Benchtop (N9020-40017) in the kit. Attach the Top Trim Strip to the
top of the Front Frame Assembly.
5.
Locate the two Side Trim Strips (5041-9172) in the kit. Remove the backing and apply one
Side Trim Strip to each side of the Front Frame Assembly.
6.
Replace the four instrument bottom feet.
7.
Replace the four key locks to the bottom feet.
8.
If the analyzer has Option EXM, External Mixer, locate the 50
Ω
termination (1810-0118) that
was removed earlier. Connect the 50
Ω
termination to the EXT MIXER connector.
Verify Hardware Installation
1.
Verify the spectrum analyzer application loads and sweeps as expected.
2.
Press
System
,
Show
Hard ware
on the analyzer and verify that Option N9030A-MTU,
Multi-touch user interface, is installed.