Installation Note N9030-90098
43
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
11.
Refer to
Figure 30
. Connect the ribbon to the BBIQ Interface Board.
12.
Install two N9020-20017 Hex screws
(1)
to secure the flex ribbon into the connector. Torque
to 4 in-lbs with the ¼-inch socket.
Figure 30
Inverter Ribbon Cable Installation