36
Installation Note N9030-90098
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Add EXT MIXER Connector and Cable to Front Frame Assembly
1.
Locate the SMA female to SMA female connector (1250-1666), ¼” lock washer (2190-0067),
and ¼” hex nut (2950-0223) that were removed from the original front frame assembly.
2.
Insert the SMA female to SMA female connector in the hole in the front frame assembly from
the front of the assembly. The hex feature on the connector should engage with the recess in
the front frame assembly. Refer to
Figure 22
. Secure the connector using the ¼” lock washer
and ¼” hex nut from the rear. Torque to 21 inch-pounds.
Figure 22
Attaching SMA Female Connector
3.
Locate the External Mixing, Front Panel semi-rigid coax assembly that was removed from the
original front frame assembly. This is W29. Note that this cable is symmetrical; either end can
be connected to the front-panel connector.
This procedure applies only to analyzers with Option EXM, External Mixer. Proceed to the next
section if Option EXM is not installed.