Installation Note N9030-90098
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Backup Calibration Data
1.
With the instrument turned on and the signal analyzer application running as normal, connect
a USB mouse to one of the instrument USB ports.
2.
Locate a USB storage device and insert it into one of the available instrument USB ports.
(Perform a virus scan on this device before use)
3.
On the front panel of the instrument, press
System
,
Alignments
,
Backup or Restore Align
Data...
. Click on
OK
. The signal analyzer application will close and the Alignment Data Wizard
will start. This might take a few minutes.
4.
The Alignment Data Wizard window as shown in
Figure 1
will appear.
Figure 1
Alignment Data Wizard
5.
Follow the on-screen instruction to
Backup
the calibration data to the USB memory device.