Installation Note N9030-90098
21
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Data Restore
1.
With the instrument turned on and the signal analyzer application running as normal, connect
a USB mouse to one of the instrument USB ports.
2.
Connect the USB memory device that was earlier used to backup the instrument calibration
data to one of the instrument USB ports.
3.
Press
System
,
Alignments
,
Backup or Restore Align Data
.
4.
The Alignment Data Wizard window as shown in
Figure 1
will appear.
5.
Follow the on-screen instruction to
Restore
the calibration data from the USB memory device.
No matter how the instrument calibration data was backed up prior to this upgrade, any
further backing up of the data will now be done through the instrument's Alignment Data
Wizard.