Installation Note N9030-90098
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Adjustments, and Performance Verification
Adjustments and performance verification testing requires the use of the calibration software. The
latest software information and downloads are available at:
http://www.keysight.com/find/calibrationsoftware
Adjustments Required
All (if the Calibration data file could not be backed-up and restored).
None (if the Calibration data file was backed-up and restored).
Performance Testing Required
All (if the Calibration data file could not be backed-up and restored).
None (if the Calibration data file was backed-up and restored).
For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your
local Keysight office access the following URL, or if in the United States, call the following
telephone number:
http://www.keysight.com/find/assist
1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)
The X-Series Signal Analyzer Calibration Files are located on the disk drive. Therefore,
replacement of the disk drive will invalidate the current instrument calibration if the original
calibration file has not been backed-up and restored. However, a full calibration is required to
assure that the instrument meets all specifications. Arrangements regarding the level of
calibration must be made between the end user and the calibration provider