8
Installation Note N9030-90098
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Installation Procedure
Analyzer Information
1.
Connect a power cord to the analyzer and turn on the analyzer.
2.
After the analyzer has completed turning on, press
System
,
Show
,
System
. Make note of the
following information from the Show System screen:
Product Number: _____________
Serial Number: ______________
Instrument S/W Revision: ______________
3.
Check for the presence of one of the following options listed below in the Show System
screen. Put a check mark or “X” after the option listed below that appears in the Show System
screen.
N9030A-PC2_____
N9030A-PC4_____
N9030A-PC6_____
N9030A-EXM_____
N9030A-BBA_____
4.
Refer to the data in
step 2
above. If the Product Number is not N9030A,
do not proceed
with
the installation of this kit. This kit is to be installed only on N9030A signal analyzers.
5.
Refer to the data in
step 3
above. If upgrade product N9030A-MTU is being installed and the
currently installed CPU is either N9030A-PC2 or N9030A-PC4,
do not proceed
with the
installation of this kit. This kit is to be installed only on N9030A signal analyzers with
N9030A-PC6 CPUs.
6.
Refer to the data in
step 3
above. If the upgrade product N9030A-MTP is being installed and
the currently installed CPU is N9030A-PC6,
do not proceed
with the installation of this kit. This
kit is to be installed only on N9030A signal analyzers with N9030A-PC2 or N9030A-PC4 CPUs.
7.
Refer to the data in
step 3
above. If either option N9030A-EXM or N9030A-BBA is installed,
there will be additional steps required to remove and replace the front panel assembly.