Operating Information
2
N432A User’s Guide
73
Editing/generating sensor calibration tables
To help achieve the best accuracy in your measurement, you can enter the
values supplied for the sensors you are using by editing the installed
sensor calibration tables or by generating your own custom tables.
You cannot delete any of the 20 existing calibration tables or create any
additional tables. However, you can edit or delete the contents of each
table. If you need another table, you should edit and rename one of the
tables. Each calibration table can contain a maximum of 80 frequency
points.
To view the calibration tables currently stored in the N432A, press
>
>
. The
Sensor Tbls
screen is displayed as
shown in
N O T E
•
If the measurement frequency does not correspond directly to a frequency in the sensor
calibration table, the N432A calculates the calibration factor using linear interpolation.
•
If you enter a frequency outside the frequency range defined in the sensor calibration
table, the N432A uses the highest or lowest frequency point in the sensor calibration
table to set the calibration factor.
•
When the
display mode is chosen, the frequency you entered and the
sensor table indicator will be displayed in the upper window. Also, pressing
>
displays the frequency you entered and the calibration factor derived from the
selected sensor table.
Single Numeric
Offsets
Tables
Sensor Cal Tables
Summary of Contents for N432A
Page 2: ......
Page 3: ......
Page 20: ...XX N432A User s Guide ...
Page 22: ...XXII N432A User s Guide ...
Page 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Page 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Page 118: ...2 Operating Information 96 N432A User s Guide ...
Page 140: ...3 Maintenance 118 N432A User s Guide ...
Page 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Page 153: ......