Operating Information
2
N432A User’s Guide
35
Setting the Measurement Frequency
Entering the frequency of the RF signal you are measuring optimizes the
accuracy and minimizes measurement uncertainty, especially when making
comparative measurements between signals.
The frequency value field is only enabled when either the
frequency- dependent offset table or sensor calibration table, or both, is
turned on. For more information on these tables, refer to
and
“Using the Calibration/Correction Factors” on
.
Procedure
Use the following steps to set the measurement frequency.
1
Press
to access the channel setup screen.
2
Press
, then use the up/down arrow key to highlight the
FDO Table
or
CF Table
setting field and press
to display the tables.
3
Use the up/down arrow key to highlight one of the table titles and
press
to highlight
.
4
Press
to complete the selection of the offset table and press
again to return to the channel setup screen. Press
>
to complete the selection of the calibration table and return to
the channel setup screen.
Offsets
Table
On
N O T E
When no data is contained in the highlighted table, the
key is disabled (grayed out).
Table
Done
Summary of Contents for N432A
Page 2: ......
Page 3: ......
Page 20: ...XX N432A User s Guide ...
Page 22: ...XXII N432A User s Guide ...
Page 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Page 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Page 118: ...2 Operating Information 96 N432A User s Guide ...
Page 140: ...3 Maintenance 118 N432A User s Guide ...
Page 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Page 153: ......