3
Maintenance
102
N432A User’s Guide
Fan
This test confirms that the internal cooling fan is running.
Real Time Clock (RTC) Battery
The RTC battery provides power for the real- time clock circuitry on the
motherboard when the N432A is powered off.
During the RTC battery test, the RTC battery power level is determined by
converting the ADC value of the RTC battery power level read from the
FPGA, to voltage level. If the battery power level is less than a
prespecified threshold (threshold to be defined later), the test will fail. An
error will be logged in the error list if the test fails.
V
RF
, V
COMP
, V
0
Paths
An array of tests on the voltage metering for V
RF
, V
COMP
, and V
0
. The
built- in self- test circuit will output a set of known DC voltages to the
bridge which are then compared to those measured by the N432A
metering. This is a user- invoked self- test.
Summary of Contents for N432A
Page 2: ......
Page 3: ......
Page 20: ...XX N432A User s Guide ...
Page 22: ...XXII N432A User s Guide ...
Page 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Page 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Page 118: ...2 Operating Information 96 N432A User s Guide ...
Page 140: ...3 Maintenance 118 N432A User s Guide ...
Page 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Page 153: ......