3
Maintenance
100
N432A User’s Guide
When the self- test has completed, press
to return to the
Self Test
menu.
If the self- test failed, information about the failure will be displayed on
the screen.
For any currently selected bridge resistance value of either 100
Ω
, 200
Ω
,
300
Ω
, or 400
Ω
, the internal self- test resistance value of 200
Ω
will be
used to perform the self- test.
When you press
followed by
even if the sensor is still
connected, you will be prompted again to disconnect the sensor. If you
cancel the sensor disconnect confirmation with the sensor still connected,
tests for V
RF
, V
COMP
, and V
0
will be skipped and displayed as shown
below.
Figure 3-2
V
RF,
V
COMP
, and V
0
tests skipped
Done
Run Self Test
Confirm