Maintenance
3
N432A User’s Guide
107
–
148
Character data not allowed
A discrete parameter was received but a character string or a numeric parameter was
expected. Check the list of parameters to verify that you have used a valid parameter
type.
Example: MEM:CLE CUSTOM_1
–
151
Invalid string data
An invalid string was received. Check to see if you have enclosed the character string in
single or double quotes.
Example; MEM:CLE “CUSTOM_1
–
158
String data not allowed
A character string was received but is not allowed for the command. Check the list of
parameters to verify that you have used a valid parameter type.
Example: LIM:STAT ‘ON’
–
161
Invalid block data
A block data element was expected but was invalid for some reason, for example,
*DDT #15FET. The 5 in the string indicates that five characters should follow, whereas in
this example there are only three.
–
168
Block data not allowed
A legal block data element was encountered but not allowed by the N432A at this point
Example: SYST:COMM:GPIB:ADDR #15FETC?
–
178
Expression data not allowed
A legal expression data was encountered but not allowed by the N432A at this point
Example: FORM:BORD (5+2)
–
211
Trigger ignored
Indicates that <GET> or *TRG, or TRIG:IMM was received and recognized by the device
but was ignored because the N432A was not in the wait-for-trigger state
Summary of Contents for N432A
Page 2: ......
Page 3: ......
Page 20: ...XX N432A User s Guide ...
Page 22: ...XXII N432A User s Guide ...
Page 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Page 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Page 118: ...2 Operating Information 96 N432A User s Guide ...
Page 140: ...3 Maintenance 118 N432A User s Guide ...
Page 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Page 153: ......