
2
Operating Information
74
N432A User’s Guide
Table 2-4
Installed thermistor sensor models
There are also 17 sensor calibration tables named
CUSTOM_0
through
CUSTOM_16
. These tables do not contain any data when the N432A is
shipped from the factory.
Editing the sensor calibration tables require the following steps:
1
Identify and select the table you want to edit
2
Rename the table
3
Enter the frequency and calibration factor pairs
4
Save the table
Procedure
First, select the table you want to edit as follows:
1
Press
>
>
to display the
Sensor Tbls
screen.
Figure 2-41
Sensor Tbls screen
Table
Sensor model
0
DEFAULT
[1]
[1] DEFAULT is a sensor calibration table where the reference calibration factor and
calibration factors are 100%. This sensor calibration table can be used during the
performance testing of the N432A.
1
478A
2
8478B
Tables
Sensor Cal Tables
Summary of Contents for N432A
Page 2: ......
Page 3: ......
Page 20: ...XX N432A User s Guide ...
Page 22: ...XXII N432A User s Guide ...
Page 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Page 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Page 118: ...2 Operating Information 96 N432A User s Guide ...
Page 140: ...3 Maintenance 118 N432A User s Guide ...
Page 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Page 153: ......