Installing N108xA on the 86100D
4
N108xA-Series Getting Started
73
Step 13. Connect to the PNA Network Analyzer
1
In the compliance application’s
Set Up
tab, select
PNA
for the network analyzer and then click
Connect PNA/ENA
.
Figure 74
Selecting an PNA Network Analyzer
2
Go to the Keysight Connection Expert, click on the PNA listing, and copy (
CTRL-C
) the SICL address
Notice that this LAN address for the PNA has the form lan,5025[
IP address
].
Figure 75
Copy PNA’s SICL Address
3
Return to the compliance application.
4
Select
Enter full SICL address or VISA alias
and paste (
CTRL-V
) the SICL address into the field as shown in
5
Click
Identify
to confirm the connection.
6
Click
OK to
close the dialog box.
NOTE
If you already connected an ENA network analyzer, skip these steps and continue with