![Keysight Technologies Infiniium MXR Series User Manual Download Page 94](http://html1.mh-extra.com/html/keysight-technologies/infiniium-mxr-series/infiniium-mxr-series_user-manual_1962332094.webp)
94
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
6
Testing Performance
500 mV/div
±20 V
352.0 mV
52.0 mV
1 V/div
±100 V
1.602 V
102.0 mV
2 V/div
±100 V
1.702 V
202 mV
5 V/div
±100 V
2.002 V
502 mV
Channel 3
5 mV/div
±2 V
32.5 mV
2.5 mV
10 mV/div
±5 V
78.0 mV
3.0 mV
20 mV/div
±10 V
154.0 mV
4.0 mV
50 mV/div
±10 V
157.0 mV
7.0 mV
100 mV/div
±20 V
312.0 mV
12.0 mV
200 mV/div
±20 V
322.0 mV
22.0 mV
500 mV/div
±20 V
352.0 mV
52.0 mV
1 V/div
±100 V
1.602 V
102.0 mV
2 V/div
±100 V
1.702 V
202 mV
5 V/div
±100 V
2.002 V
502 mV
Channel 4
5 mV/div
±2 V
32.5 mV
2.5 mV
10 mV/div
±5 V
78.0 mV
3.0 mV
20 mV/div
±10 V
154.0 mV
4.0 mV
50 mV/div
±10 V
157.0 mV
7.0 mV
100 mV/div
±20 V
312.0 mV
12.0 mV
200 mV/div
±20 V
322.0 mV
22.0 mV
500 mV/div
±20 V
352.0 mV
52.0 mV
1 V/div
±100 V
1.602 V
102.0 mV
2 V/div
±100 V
1.702 V
202 mV
5 V/div
±100 V
2.002 V
502 mV
Channel 5
5 mV/div
±2 V
32.5 mV
2.5 mV
10 mV/div
±5 V
78.0 mV
3.0 mV
Table 8
Offset Accuracy Test (1 M
Ω
) (continued)
Vertical Scale
Ext.
Supply
Setting
V
DMM
+
V
Scop
e+
V
error
+
V
DMM
-
V
Scop
e-
V
error-
Offset
Accuracy
Limit (±)
V
zero-error
V
zero-error
Limit (±)
Summary of Contents for Infiniium MXR Series
Page 1: ...Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide...
Page 10: ...10 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide Index...
Page 20: ...20 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 1 Setting Up...
Page 106: ...106 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 6 Testing Performance...