Testing Performance
6
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
85
20
Repeat step 15 through step 19 to complete measuring gains for remaining
scale settings for channel 1 in
of the Performance Test Record.
21
For each of the remaining channels, move the splitter to the channel under test
and change the oscilloscope configuration as follows:
a
Click
Control > Factory Default
.
b
Ensure the channel under test is displayed and all other channels are turned
off.
c
Set the vertical scale of the channel to 5 mV/div.
d
Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz
waveform).
e
Click
Trigger> Setup Trigger...
and change the source to the channel.
f
Click
Measure > Add Measurement...
. Select the Vertical RMS voltage
measurement, the channel under test as the source, Entire Display as the
Measurement Area, and AC for the RMS Type.
g
Repeat step 6d and step 7 to step 20 to complete measuring gains for the
channel.