Procedure
1
Preset the power meter.
2
Ensure that the power sensor is disconnected from any source and zero the
meter.
3
Connect the power sensor to the power meter's Power Ref connector and
calibrate the meter.
4
Make the connections to oscilloscope channel 1 as shown in the preceding
connection diagram.
5
Set up the power meter to display measurements in units of Watts.
6
Click
Control > Factory Default
, then configure the oscilloscope as follows:
a
Ensure the channel under test is displayed and all other channels are turned
off.
b
In the Channel dialog box, set the vertical scale of the channel under test to
5 mV/div and the input impedance to 50 Ohms.
c
Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz
waveform).
d
Click
Setup > Acquisition...
and ensure the acquisition parameters are set up
as follows:
•
Sin(x)/x Interpolation = Auto
•
Analog Averaging = Enabled with # of Averages set to 16
•
Analog Memory Depth = Automatic
•
Analog Sampling rate = Maximum (select the Manual button and click
the up-arrow as high as it goes)
e
Click
Measure > Add Measurement...
.
f
In the Add Measurement dialog box, click Vertical and RMS, and configure
the measurement as follows:
•
Source = Channel 1
•
Measurement Area = Entire Display
•
RMS Type = AC
NOTE
•
Connect output 1 of the 11667B splitter to the oscilloscope Channel n input directly using the
1250-1200 adapter, without any additional cabling or adapters.
•
Connect output 2 of the 11667B power splitter to the power sensor using the 1250-1750 N-type to
3.5 mm adapter, without any additional cabling or adapters.
•
Minimize the use of other adapters.
•
Ensure that SMA and 3.5 mm connectors are tightened properly:
8 in-lbs (90 N-cm) for 3.5 mm
5 in-lbs (56 N-cm) for SMA
Summary of Contents for Infiniium MXR Series
Page 1: ...Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide...
Page 10: ...10 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide Index...
Page 20: ...20 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 1 Setting Up...
Page 106: ...106 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 6 Testing Performance...