![Keysight Technologies Infiniium MXR Series User Manual Download Page 92](http://html1.mh-extra.com/html/keysight-technologies/infiniium-mxr-series/infiniium-mxr-series_user-manual_1962332092.webp)
92
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
6
Testing Performance
Channel 5
5 mV/div
±60 mV
3.1 mV
2.5 mV
10 mV/div
±120 mV
4.2 mV
3.0 mV
20 mV/div
±240 mV
6.4 mV
4.0 mV
50 mV/div
±600 mV
13.0 mV
7.0 mV
100 mV/div
±1.2 V
24.0 mV
12.0 mV
200 mV/div
±2.4 V
58.0 mV
22.0 mV
500 mV/div
±4.0 V
112.0 mV
52.0 mV
1 V/div
±4.0 V
162.0 mV
102.0 mV
Channel 6
5 mV/div
±60 mV
3.1 mV
2.5 mV
10 mV/div
±120 mV
4.2 mV
3.0 mV
20 mV/div
±240 mV
6.4 mV
4.0 mV
50 mV/div
±600 mV
13.0 mV
7.0 mV
100 mV/div
±1.2 V
24.0 mV
12.0 mV
200 mV/div
±2.4 V
58.0 mV
22.0 mV
500 mV/div
±4.0 V
112.0 mV
52.0 mV
1 V/div
±4.0 V
162.0 mV
102.0 mV
Channel 7
5 mV/div
±60 mV
3.1 mV
2.5 mV
10 mV/div
±120 mV
4.2 mV
3.0 mV
20 mV/div
±240 mV
6.4 mV
4.0 mV
50 mV/div
±600 mV
13.0 mV
7.0 mV
100 mV/div
±1.2 V
24.0 mV
12.0 mV
200 mV/div
±2.4 V
58.0 mV
22.0 mV
500 mV/div
±4.0 V
112.0 mV
52.0 mV
1 V/div
±4.0 V
162.0 mV
102.0 mV
Channel 8
5 mV/div
±60 mV
3.1 mV
2.5 mV
Table 7
Offset Accuracy Test (50
Ω
) (continued)
Vertical Scale
Ext.
Supply
Setting
V
DMM
+
V
Scop
e+
V
error
+
V
DMM
-
V
Scop
e-
V
error-
Offset
Accuracy
Limit (±)
V
zero-error
V
zero-error
Limit (±)
Summary of Contents for Infiniium MXR Series
Page 1: ...Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide...
Page 10: ...10 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide Index...
Page 20: ...20 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 1 Setting Up...
Page 106: ...106 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 6 Testing Performance...