Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
6
Testing Performance
18
Using the Channel dialog box, set the channel's offset value to 60.0 mV and the
impedance to 50
Ω
(or to an offset of 2 V and the impedance to 1 M
Ω
if this is
your second pass through this test and you are using the 1 M
Ω
input
impedance version).
19
Set the DC supply voltage (External Supply Setting) to +60.0 mV (or +2 V for
the 1 M
Ω
version of this test).
20
Press
[Clear Display]
on the oscilloscope, wait for the number of averages to
return to 256, and then record the DMM voltage reading as VDMM+ and the
scope V avg reading as in either
(depending on
which version of the test you are on). Fill in the column by using the
following equation:
= VDMM+ -
must be within the limits specified by the corresponding Offset
Accuracy Limit listed in the table.
21
Change the channel 1 offset value to -60.0 mV (or -2 V for the 1 M
Ω
version of
this test).
22
Set the DC supply voltage to -60.0 mV (or -2 V for the 1 M
Ω
version of this
test).
23
Press
[Clear Display]
on the oscilloscope, wait for the number of averages to
return to 256, and then record the DMM voltage reading as VDMM- and the
scope V avg reading as VScope- in either
or
(depending on
Summary of Contents for Infiniium MXR Series
Page 1: ...Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide...
Page 10: ...10 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide Index...
Page 20: ...20 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 1 Setting Up...
Page 106: ...106 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 6 Testing Performance...