72
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
6
Testing Performance
6
Change the vertical scale of channel 1 to 10 mV/div, press
[Clear Display]
, wait
for the number of averages to return to 256, and then record the oscilloscope's
mean V avg reading in the Vzero-error column of
. Use the Channel
dialog box to select 1 M
Ω
impedance and repeat for 1 M
Ω
. Record the results in
7
Repeat step 6 for the remaining vertical scale settings for channel 1 in
of the Performance Test Record.
8
Click
Control > Factory Default
, turn off channel 1, and turn the channel 2 display
on.
9
Configure the oscilloscope to measure V avg on channel 2 as follows:
a
In the Acquisition dialog box, enable averaging and set # of Averages to 256.
b
Change the vertical scale of channel 2 to 5 mV/div.
c
Drag and drop the Average voltage measurement icon onto the channel 2
waveform.
10
Press
[Clear Display]
on the oscilloscope, wait for the number of averages to
return to 256, and then record the oscilloscope's mean V avg reading in the
. Use the Channel dialog box to select 1 M
Ω
impedance and repeat for 1 M
Ω
Summary of Contents for Infiniium MXR Series
Page 1: ...Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide...
Page 10: ...10 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide Index...
Page 20: ...20 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 1 Setting Up...
Page 106: ...106 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 6 Testing Performance...