Testing Performance
6
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
71
5
Press
[Clear Display]
on the oscilloscope and wait for the number of averages
display (top right area of screen) to return to 256. Record the oscilloscope's
mean V avg reading in the Vzero-error column of
of the Performance
Test Record. Use the Channel dialog box to select 1 M
Ω
impedance and then
repeat step 3 to step 5. Record the reading in
for channel 1.
NOTE
•
For all oscilloscope readings in this procedure, use the mean value in the Measurement Results
pane at the bottom of the screen.
•
If a question mark appears in front of any values in the Results pane, press
[Clear Display]
on the
oscilloscope, wait for the number of averages to return to 256, and then record the oscilloscope
reading. The question mark indicates that the oscilloscope could not make a reliable
measurement.