68
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
6
Testing Performance
8
Repeat step 3 through step 7 on the remaining channels of the oscilloscope.
9
Repeat step 3 through step 8 using an input resistance of 1 M
Ω
instead of 50
Ω
and verify resistance readings of 1 M
Ω
±10 k
Ω
.
NOTE
Due to the current source limitations of the 34411A/3458A/34465A DMM, be sure to set it to
10 M
Ω
range instead of Auto Range when testing the 1 M
Ω
resistance.