![Keysight Technologies Infiniium MXR Series User Manual Download Page 101](http://html1.mh-extra.com/html/keysight-technologies/infiniium-mxr-series/infiniium-mxr-series_user-manual_1962332101.webp)
Testing Performance
6
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
101
1 V/div
±3.0 V
±2 %
2 V/div
±6.0 V
±2 %
5 V/div
±15.0 V
±2 %
Channel 6
5 mV/div
±15 mV
±2 %
10 mV/div
±30 mV
±2 %
20 mV/div
±60 mV
±2 %
50 mV/div
±150 mV
±2 %
100 mV/div
±300 mV
±2 %
200 mV/div
±600 mV
±2 %
500 mV/div
±1.5 V
±2 %
1 V/div
±3.0 V
±2 %
2 V/div
±6.0 V
±2 %
5 V/div
±15.0 V
±2 %
Channel 7
5 mV/div
±15 mV
±2 %
10 mV/div
±30 mV
±2 %
20 mV/div
±60 mV
±2 %
50 mV/div
±150 mV
±2 %
100 mV/div
±300 mV
±2 %
200 mV/div
±600 mV
±2 %
500 mV/div
±1.5 V
±2 %
1 V/div
±3.0 V
±2 %
2 V/div
±6.0 V
±2 %
5 V/div
±15.0 V
±2 %
Channel 8
5 mV/div
±15 mV
±2 %
10 mV/div
±30 mV
±2 %
20 mV/div
±60 mV
±2 %
Table 10
DC Gain Accuracy Test (1 M
Ω
) (continued)
Vertical
Scale
Ext. Supply
Setting
V
DMM+
V
Scope+
V
DMM-
V
Scope-
Calculated
DC Gain
Error
DC Gain
Error Test
Limits
Summary of Contents for Infiniium MXR Series
Page 1: ...Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide...
Page 10: ...10 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide Index...
Page 20: ...20 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 1 Setting Up...
Page 106: ...106 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 6 Testing Performance...