85
5. Performance Verification
Bandwidth Performance Verification
NOTE
You only need to perform the Bandwidth Performance Verification test with the ZIF
tip as described above. If the probe passes with the ZIF probe head then it will pass
with the browser probe head as well. You are not required to test the bandwidth
specification with both probe heads. If you have the choice between using the ZIF
probe head or the browser, Keysight suggests using the ZIF probe head because
measurements are more repeatable with it and it is easier to make a proper
connection.
To measure the performance of the N5445A Browser with the N2803A
probe amp, the same procedure as described above is used except the
browser is held in position with a probe positioner. This setup is shown in
using the Keysight N2787A 3D Probe Positioner.
a
Tilt the browser slightly so the two tips touch the center trace and
the ground plane on the PV fixture at about the same time.
b
Probe the halfway point between the where the pinchers are and
where the microstrip transitions into the rectangular coax line.
c
Set the span of the browser to the minimum possible as should in
this picture:
Summary of Contents for InfiniiMax III Series
Page 1: ...Keysight InfiniiMax III Series Probes User s Guide...
Page 4: ...4 Contents...
Page 28: ...28 1 Using InfiniiMax III Series Probes Simplified InfiniiMax III Schematic...
Page 64: ...64 2 InfiniiMax III Probe Heads N5444A InfiniiMax III 2 92 mm 3 5 mm SMA Probe Head...
Page 100: ...100 5 Performance Verification Performance Test Record...
Page 104: ...104 6 Performance Plots N2836A 26 GHz Solder In Head with N2803A...
Page 105: ...105 6 Performance Plots N2836A 26 GHz Solder In Head with N2803A...
Page 107: ...107 6 Performance Plots N2838A N5439A 25 GHz PCB ZIF Tip Head with N2803A...
Page 108: ...108 6 Performance Plots N2838A N5439A 25 GHz PCB ZIF Tip Head with N2803A...
Page 110: ...110 6 Performance Plots N5440A N5439A 28 GHz Ceramic ZIF Tip Head with N2803A...
Page 111: ...111 6 Performance Plots N5440A N5439A 28 GHz Ceramic ZIF Tip Head with N2803A...
Page 113: ...113 6 Performance Plots N5447A N5439A 28 GHz Ceramic High Sensitivity ZIF Tip Head with N2803A...
Page 114: ...114 6 Performance Plots N5447A N5439A 28 GHz Ceramic High Sensitivity ZIF Tip Head with N2803A...
Page 116: ...116 6 Performance Plots N5441A 16 GHz Solder In Head with N2803A...
Page 117: ...117 6 Performance Plots N5441A 16 GHz Solder In Head with N2803A...
Page 119: ...119 6 Performance Plots N5444A 28 GHz SMA 2 92 mm Head with N2803A...
Page 121: ...121 6 Performance Plots N5445A 30 GHz Browser Head 1 mm span with N2803A...
Page 122: ...122 6 Performance Plots N5445A 30 GHz Browser Head 1 mm span with N2803A...