74
4. Calibration / Deskew Procedure
Procedure
12
A dialog box will then open that tells you how to connect the probe
head to the calibration/deskew fixture.
The fixture has three spring-loaded fingers (shown below) that clamp
probe head wires to the fixture if you are using the ZIF probe head or the
Solder-in probe head.
13
On the fixture, the center gold trace is signal and the large plates on
either side are both ground as shown below.
Summary of Contents for InfiniiMax III Series
Page 1: ...Keysight InfiniiMax III Series Probes User s Guide...
Page 4: ...4 Contents...
Page 28: ...28 1 Using InfiniiMax III Series Probes Simplified InfiniiMax III Schematic...
Page 64: ...64 2 InfiniiMax III Probe Heads N5444A InfiniiMax III 2 92 mm 3 5 mm SMA Probe Head...
Page 100: ...100 5 Performance Verification Performance Test Record...
Page 104: ...104 6 Performance Plots N2836A 26 GHz Solder In Head with N2803A...
Page 105: ...105 6 Performance Plots N2836A 26 GHz Solder In Head with N2803A...
Page 107: ...107 6 Performance Plots N2838A N5439A 25 GHz PCB ZIF Tip Head with N2803A...
Page 108: ...108 6 Performance Plots N2838A N5439A 25 GHz PCB ZIF Tip Head with N2803A...
Page 110: ...110 6 Performance Plots N5440A N5439A 28 GHz Ceramic ZIF Tip Head with N2803A...
Page 111: ...111 6 Performance Plots N5440A N5439A 28 GHz Ceramic ZIF Tip Head with N2803A...
Page 113: ...113 6 Performance Plots N5447A N5439A 28 GHz Ceramic High Sensitivity ZIF Tip Head with N2803A...
Page 114: ...114 6 Performance Plots N5447A N5439A 28 GHz Ceramic High Sensitivity ZIF Tip Head with N2803A...
Page 116: ...116 6 Performance Plots N5441A 16 GHz Solder In Head with N2803A...
Page 117: ...117 6 Performance Plots N5441A 16 GHz Solder In Head with N2803A...
Page 119: ...119 6 Performance Plots N5444A 28 GHz SMA 2 92 mm Head with N2803A...
Page 121: ...121 6 Performance Plots N5445A 30 GHz Browser Head 1 mm span with N2803A...
Page 122: ...122 6 Performance Plots N5445A 30 GHz Browser Head 1 mm span with N2803A...