Keysight E8740A-070 Start-up Guide
1
Connect the measurement instruments to the radar generation system as
shown in
below.
a
During calibration, IQTools requires remote control of the signal analyzer.
Connect a USB cable from the M9573A to the Signal Analyzers’ slave USB
port at the rear of the instrument.
b
Synchronize the timing base of the measurement system with the radar
signal generator system by connecting a coaxial cable from the
ref clk out
to the rear
Ext Ref In
port of the Spectrum Analyzer.
c
Connect the M1971E smart mixer to the Signal Analyzer as shown. A 10
MHz reference from the Spectrum Analyzer to the Smart Mixer is required
when using dual conversion path. Note that if the Spectrum Analyzer is
already powered up and ready, connect the coaxial cables first before the
USB cable.
Figure 18
Calibration setup using Smart Mixer and SA ext. mixer port