5-26
Keysight B1542A User’s Guide, Edition 10
PLSDIV Test Definitions
PLSDIV IdVg, PLSDIV IdVg [2]
Test Output
The following output parameters are displayed after the test.
Vgate
Target gate voltage Vg, in V.
Idrain
Drain current Id, in A.
VgMon
Gate voltage, in V. Oscilloscope monitor value.
VgConv
Convergence result of the gate voltage automatic adjustment.
Vdrain
Target drain voltage Vd, in V.
VdSet
Drain voltage, in V. SMU output value.
VdInt
Drain voltage, in V. Oscilloscope monitor value.
VdConv
Convergence result of the drain voltage automatic adjustment.
IdPerWg
Drain current per unit gate width, in A/
m.
IdPerWg = Id / Wg
VgMon, VdInt, and Id values will be the value obtained at the data extraction timing
on the pulse top. See
.
VdConv and VgConv returns 0 (convergence condition is satisfied) or
-
1 (not
satisfied). If the automatic adjustment is OFF, 0 is returned.
If the drain voltage automatic adjustment function is ON, VdInt will be close to Vd.
And if the function is OFF, VdSet will be close to Vd.
If the gate voltage automatic adjustment function is ON, VgMon will be close to Vg.
And if the function is OFF, Vpgu-out will be close to Vg.
Figure 5-5
Test Output Parameters
D
S
G
Shunt resistor, Z
VdSet
Id
VgMon
Vpgu-out
VdInt = VdSet Z Id
PGU
SMU
Sub
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......