5-30
Keysight B1542A User’s Guide, Edition 10
PLSDIV Test Definitions
PLSDIV IV SMU
Table 5-2
Integration Time Setup Parameters for E5260/E5270
ADCType
IntegMode
IntegNum
HighSpeed
Auto
Value that defines the number of averaging samples given by the
following formula. 1 to 1023.
Number of averaging samples
=
IntegNum
system-defined-value
where
system-defined-value
is the number of averaging samples
automatically set by the E5260/E5270.
Manual
Number of averaging samples. 1 to 1023.
PLC
Value that defines the number of averaging samples given by the
following formula. 1 to 100.
Number of averaging samples
=
IntegNum
128
The E5260/E5270 gets 128 samples in a power line cycle, repeats
this for the times you specify, and performs averaging to get the
measurement data.
HighResolution
Auto
Value that defines the integration time given by the following
formula. 1 to 127.
Integration time
=
IntegNum
system-defined-value
where
system-defined-value
is the integration time automatically
set by the E5260/E5270.
Manual
Value that defines the integration time given by the following
formula. 1 to 127.
Integration time
=
IntegNum
80
s
PLC
Value that defines the integration time given by the following
formula. 1 to 100.
Integration time
=
IntegNum
/
power line frequency
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......