5-38
Keysight B1542A User’s Guide, Edition 10
PLSDIV Test Definitions
Parameters
VdAdjust
VdAdjust sets the drain voltage automatic adjustment function ON or OFF. This
function is used to compensate the voltage drop caused by the shunt resistor. See
. When this function is OFF, the SMU applies Vd (specified drain
voltage). When this function is ON, the SMU output level will be automatically
controlled so that Vd is applied to the drain terminal.
If you do not mind the effect of the shunt resistor, set this function OFF. This brings
higher measurement throughput.
VgAdjust
VgAdjust sets the gate voltage automatic adjustment function ON or OFF. This
function is used to obtain stable gate voltage. When this function is OFF, the PGU
applies Vg (specified gate voltage). When this function is ON, the PGU output level
will be automatically controlled so that Vg is applied to the gate terminal.
If you do not mind the stability of the gate voltage, set this function OFF. This
brings higher measurement throughput.
If you set PulseBase < > 0 V, set this function ON.
To use this function, set the pulse amplitude to < 4.0 V.
Figure 5-11
VdAdjust and VgAdjust
D
S
G
Shunt resistor, Z
Vd (set)
Vd (monitored) = Vd (set) Z Id
Id
Vg (monitored)
Vg (set)
PGU
SMU
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......