3-12
Keysight B1542A User’s Guide, Edition 10
Performing System Setup and Compensation
Executing Action
Executing Action
Execute the action as shown below. Before starting the action, confirm that all
system components are connected to the system controller and available on the same
GPIB bus and on the same LAN for the DSO/MSO 9000 series or S series
oscilloscope.
1. Open the PLSDIV Setup test setup from your preset group.
2. Set the DcSwitch field to Yes. Do not set No.
3. Click the Extended Setup button and set the SWAddress value.
SWAddress
GPIB address of the switch controller. Integer, 1 to 32.
4. Select the action by using the Action field.
•
System Configuration
“System Configuration” on page 3-10
•
Compensation
“Compensation Action” on page 3-13
•
Skew Measurement
•
Drain Cable Replacement
•
Pgu Compensation
“Pgu Compensation Action” on page 3-13
.
5. Click the Single button at the upper right corner of the EasyEXPERT screen or
the Desktop EasyEXPERT screen.
6. Follow the pop-up window and dialog box to complete the action.
You will get navigation for performing the system setup and completing the
action.
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......