3-10
Keysight B1542A User’s Guide, Edition 10
Performing System Setup and Compensation
System Configuration
System Configuration
The system configuration must be set after completing the system installation or
after changing a channel assignment or after changing the interface address of the
system components.
Procedure:
1. Open the PLSDIV Setup test setup from your preset group.
2. Set the Action field to
System Configuration
.
3. Click the Single button at the upper right corner on the EasyEXPERT screen or
the Desktop EasyEXPERT screen. The Configuration dialog box is opened.
4. Set and save the system configuration parameters by using the Configuration
dialog box. For the setup parameters, see
LAN Setup for DSO Control Dialog Box
This dialog box is opened by clicking the Option button on the Configuration dialog
box, and is used to set the IP address of the DSO/MSO 9000 series or S series
oscilloscope. Set the IP address properly and click OK.
Table 3-5
LAN Setup for DSO Control Dialog Box
Field Name
Description
IP Address
Enter the IP address of the DSO/MSO 9000 series or S
series oscilloscope.
Device Name
This field always displays
inst0
which must be the
device name of the oscilloscope for the pulsed IV test
system.
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......