Keysight B1542A User’s Guide, Edition 10
2-5
Installation
DC Probes
DC Probes
The MOSFET contact pads for DC measurement shown in
, are more
popular than the RF contact pads shown in
. If device under test is
configured with DC contact pads, use DC probes instead of RF probes. The DC
probes are better suited for contact with the DC contact pads than the RF probes.
See
for the contact pads and the DC probes. Required cables for
connecting the DC probes are furnished with the B1542A as the DC probe cable set.
For more information, see
“Connecting DC Probes” on page 2-14
When using DC probes, minimum pulse width should be limited to approximately
60 ns to 100 ns.
Figure 2-3
Contact Pad and DC Probe Connection
Terminator
Well
Gate
Source
Drain
This makes a current return path
for the drain current signal.
This makes a current return path
for the gate pulse signal.
This shorts Well and Source.
to Test system
to Test system
DC probes
DC probes
Signal
Signal
Gnd
Gnd
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......